User`s guide

Agilent B2900 Users Guide, Edition 2 3- 15
Installation
Connecting a DUT
CAUTION Maximum voltage and current
The test fixture and adapters must be used under the following limitations to prevent
damage.
Agilent N1294A-001: 250 V maximum, 42 V maximum for connecting N1295A
Agilent N1294A-002: 250 V maximum, 42 V maximum for connecting N1295A
Agilent N1295A: 42 V, 1.05 A maximum
Using the 16442B Test Fixture
Agilent 16442B is a test fixture which has six triaxial connectors for source/measure
unit (SMU), the GNDU connector for the ground unit of a parameter/device
analyzer, the Intlk connector for the interlock control, and six coaxial connectors for
other instruments. See Agilent 16442B Users Guide for details and accessories of
the 16442B.
Requirements Agilent 16442B test fixture furnished with pin clip wire, pin plug wire, etc.
Agilent N1294A-001 banana to triaxial adapter for 2-wire connections
Agilent N1294A-002 banana to triaxial adapter for 4-wire connections
Triaxial cable, 2 ea. for 2-wire connections, and 3 ea. for 4-wire connections
NOTE Shielding
16442B is equipped with a lid. To minimize the effect of ambient noise, close the lid
when performing measurements.
CAUTION Maximum voltage and current
The test fixture and adapters must be used under the following limitations to prevent
damage.
Agilent N1294A-001: 250 V maximum, 200 V maximum for connecting 16442B
Agilent N1294A-002: 250 V maximum, 200 V maximum for connecting 16442B
Agilent 16442B: 200 V, 1 A maximum for SMU input