User`s guide

Agilent B2900 Users Guide, Edition 2 2- 27
Introduction
Specifications
General Specifications
Environment
For use in indoor facilities
Temperature range
Operating 0 C to +55 C
Storage -30 C to +70 C
Humidity range
Operating 30 % to 80 %RH, non-condensing
Storage 10 % to 90 %RH, non-condensing
Altitude
Operating 0 m to 2,000 m (6,561 ft.)
Storage 0 m to 4,600 m (15,092 ft.)
Power requirement: 90 V to 264 V, 47 Hz to 63 Hz
Maximum volt-amps (VA): 250 VA
Cooling
Forced air. Side intake and rear exhaust.
Regulatory compliance
EMC IEC61326-1 / EN61326-1
AS/NZS CISPR 11
Safety IEC61010-1 / EN61010-1
CAN/CSA-C22.2 No. 61010-1-04, C/US
Certifications: CE, cCSAus, C-Tick
•Dimensions
Without handle and feet: 88 mm (H) 213 mm (W) 450 mm (D)
Bench configuration (with handle and feet): 180 mm (H) 260 mm (W) 480
mm (D)
•Weight
B2901A/B2911A: 5.0 kg
B2902A/B2912A: 6.4 kg