User`s guide

2- 24 Agilent B2900 Users Guide, Edition 2
Introduction
Specifications
High Capacitance Mode
The high capacitance mode permits the measurement of devices with capacitances
greater than the normal mode maximum load value of 0.01 F. In high capacitance
mode, the maximum allowed value of the load capacitance is 50 F.
Voltage output settling time:
Time required to reach within 0.1 % of final value with 4.7 F capacitive load
on a fixed range at specified current limit value.
200 mV and 2 V Ranges: 600 s, at 1 A limit
20 V Range: 1.5 ms, at 1 A limit
200 V Range: 20 ms, at 100 mA limit
Current measurement settling time:
Time required to reach within 0.1 % of final value after voltage source is
stabilized on fixed range. Vout is 5 V unless noted.
1 A range: 230 ms
10 A and 100 A ranges: 23 ms
1 mA and 10 mA ranges: 0.23 ms
100 mA to 3 A ranges: 100 s
Mode change delay:
Delay into High Capacitance mode:
1 A range: 230 ms
10 A and 100 A ranges: 23 ms
1 mA to 3 A ranges: 1 ms
Delay out of High Capacitance mode:
All ranges: 10 ms
Noise 10 Hz to 20 MHz (20 V range): 4.5 mVrms
Voltage source range change overshoot (20 V range or below):
< 250 mV, 20 MHz bandwidth