Agilent B2900 User’s Guide, Edition 2 2- 23
Introduction
Specifications
Other Supplemental Characteristics
Output Characteristics
• Sensing modes: 2-wire or 4-wire (remote-sensing) connections
• Low terminal connection: Chassis grounded or floating
• Output connectors:
Banana jack. Triaxial connections are recommended for sourcing and measuring
less than 1 nA. A banana jack to triaxial adapter is available for low current
measurement.
• Output location: Channel 1 at front and channel 2 at rear
•Maximum load:
• Normal mode: 0.01 F
• High capacitance mode: 50 F
• DC floating voltage: Maximum 250 V DC between Low Force and chassis
ground
• Guard offset voltage (V source): < 4 mV
• Remote sense operation range:
• Maximum voltage between High Force and High Sense: 3 V
• Maximum voltage between Low Force and Low Sense: 3 V
• Common mode isolation: > 1 G, < 4500 pF
• Maximum sense lead resistance: 1 k for rated accuracy
• Sense input impedance: > 10 G