User`s guide
2- 22 Agilent B2900 User’s Guide, Edition 2
Introduction
Specifications
Table 2-12 Maximum Duty Cycle and Pulse Width
Table 2-13 Minimum Pulse Width at the Given Voltage, Current, and Settling Conditions
Table 2-14 Derating Accuracy with PLC Setting < 1 PLC
Maximum
duty cycle
Pulse width
Maximum
voltage
Maximum peak
current
Maximum base
current
99.9999 % 50 s to 99999.9 s 6 V
3.03 A
a
a. Max current limitation: For 21 V/1.515 A and 6 V/3.03 A ranges, total max current is limited as
shown in
Table 2-4 for using 2 channels. Max current is not limited for using 1 channel only.
3.03 A
a
21 V
1.515 A
a
1.515 A
a
210 V 0.105 A 0.105 A
2.5 % 50 s to 1 ms 6 V 10.5 A 500 mA
50 s to 10 ms 180 V 1.05 A 50 mA
50 s to 2.5 ms 200 V 1.515 A 50 mA
Source value Limit value Load
Source settling
(%
of range)
Minimum pulse
width
200 V 1.5 A 200 0.1 % 1 ms
6 V 10.5 A 0.6 0.1 % 0.2 ms
1.5 A 200 V 65 0.1 % 2.5 ms
10.5 A 6 V 0.5 0.1 % 0.2 ms
PLC setting
Voltage range Current range
0.2 V 2 V to 200 V 10 nA 100 nA 1 A to 100 mA 1 A to 3 A
0.1 PLC 0.01 % 0.01 % 0.1 % 0.01 % 0.01 % 0.01 %
0.01 PLC 0.05 % 0.02 % 1 % 0.1 % 0.05 % 0.02 %
0.001 PLC 0.5 % 0.2 % 5 % 1 % 0.5 % 0.2 %