User`s guide

Agilent B2900 Users Guide, Edition 2 2- 21
Introduction
Specifications
Measurement Supplemental Characteristics
Temperature coefficient (0 C to 18 C and 28 C to 50 C):
(0.1 Accuracy specification) / C
•Over range:
102 % of measurement range for 1.5 A and 3 A ranges. 106 % of measurement
range other than 1.5 A and 3 A ranges.
Voltage measurement range change overshoot:
< 250 mV. 100 k load, 20 MHz bandwidth
Current measurement range change overshoot:
< 250 mV/R load, 20 MHz bandwidth
Derating accuracy for measurement speed less than 1 PLC:
Add % of range using Table 2-14 for measurement with PLC < 1.
Triggering and Timing Specifications
Triggering (Supplemental Characteristics)
Digital I/O trigger in to trigger out: 5 s
Digital I/O trigger in to source change: 5 s
LXI trigger in to source change: 200 s
Timer
Trigger timing resolution: 1 s to 100 ms
Timer accuracy: 50 ppm
Arm/trigger delay: 0 to 100000 s
Arm/trigger interval:
B2901A/B2902A: 20 s to 100000 s
B2911A/B2912A: 10 s to 100000 s
Arm/trigger event: 1 to 100000
Time stamp: TIMER value automatically saved when each measurement is
triggered.