User`s guide

Agilent B2900 Users Guide, Edition 2 2- 19
Introduction
Specifications
Source Supplemental Characteristics
Temperature coefficient (0 C to 18 C and 28 C to 50 C):
(0.1 Accuracy specification) / C
Maximum output power and source/sink limits:
31.8 W. 6 V@ 3.03 A, 21 V@ 1.515 A, 210 V@ 105 mA, four quadrant
source or sink operation.
Current limit/compliance:
Accuracy is same as current source. Minimum value is 1 % of range, or 1 nA in
10 nA range.
Voltage limit/compliance:
Accuracy is same as voltage source. Minimum value is 1 % of range, or 20 mV
in 200 mV range.
•Over range:
101 % of source range for 1.5 A and 3 A ranges. 105 % of source range other
than 1.5 A and 3 A ranges. No over range for 200 V range with current
exceeding 105 mA pulse only condition.
Over temperature protection:
Turns off all channel outputs and locks the causal channel at over temperature
sensed internally.
Voltage output settling time:
Time required to reach within 0.1 % of final value at open load condition. Step is
+10 % to +90 % range.
200 mV and 2 V ranges: < 50 s
20 V range: < 110 s
200 V range: < 700 s