User`s guide
Agilent B2900 User’s Guide, Edition 2 2- 17
Introduction
Specifications
Table 2-8 Voltage Source Specifications
Table 2-9 Current Source Specifications
Range
Programming resolution Accuracy
(gain error (% of
reading) + offset
error)
Noise
a
(peak-peak)
0.1
Hz to 10 Hz
a. Supplemental characteristics
Maximum
voltage
(over range)
B2901A
B2902A
B2911A
B2912A
200 mV 1 V 0.1 V (0.015 % + 225 V) 10 V 210 mV
2 V 10 V 1 V (0.02 % + 350 V) 20 V 2.1 V
20 V 100 V 10 V (0.015 % + 5 mV) 200 V 21 V
200 V 1 mV 100 V (0.015 % + 50 mV) 2 mV 210 V
Range
Programming resolution Accuracy
(gain error (% of
reading) + offset
error)
Noise
a
(peak-peak)
0.1
Hz to 10 Hz
a. Supplemental characteristics
Maximum
current
(over range)
B2901A
B2902A
B2911A
B2912A
10 nA
b
b. 10 nA range is not available for B2901A/B2902A.
―
10 fA (0.10 % + 50 pA) 1 pA 10.5 nA
100 nA 1 pA 100 fA (0.06 % + 100 pA) 2 pA 105 nA
1 A 10 pA 1 pA (0.025 % + 500 pA) 25 pA 1.05 A
10 A 100 pA 10 pA (0.025 % + 1.5 nA) 60 pA 10.5 A
100 A 1 nA 100 pA (0.02 % + 25 nA) 2 nA 105 A
1 mA 10 nA 1 nA (0.02 % + 200 nA) 6 nA 1.05 mA
10 mA 100 nA 10 nA (0.02 % + 2.5 V) 200 nA 10.5 mA
100 mA 1 A 100 nA (0.02 % + 20 V) 600 nA 105 mA
1 A 10 A 1 A (0.03 % + 1.5 mA) 70 A 1.05 A
1.5 A 10 A 1 A (0.05 % + 3.5 mA) 100 A 1.515 A
3 A 100 A 10 A (0.4 % + 7 mA) 120 A 3.03 A
10 A
c
c. 10 A range is available only for pulse mode, not available for DC mode.
100 A 10 A
(0.4 % + 25 mA)
d
d. Measurement speed: 0.01 PLC
―
10.5 A