User`s guide
Agilent B2900 User’s Guide, Edition 2 6- 31
Function Details
Trigger System
Synchronous Device Actions
This section applies only to when performing synchronous channel operation on a
2-channel model.
If channels are set as shown below, the device actions start simultaneously.
• To synchronize transient actions (source output)
• Trigger source is set to the same mode.
• Delay time is set to the same value.
• Source output ranging mode is set to the fixed mode.
• Source wait time control is set to OFF.
• Measurement wait time control is set to OFF.
• Measurement ranging mode is set to the fixed mode.
• To synchronize acquire actions (measurement)
• Trigger source is set to the same mode.
• Delay time is set to the same value.
• Measurement wait time control is set to OFF.
• Measurement ranging mode is set to the fixed mode.