User`s guide
Agilent B2900 User’s Guide, Edition 2 6- 29
Function Details
Trigger System
Device Actions
B2900 supports the following device actions.
• Triggered transient (source) action
After the transient trigger condition is satisfied, the channel applies a new
voltage or current value after the source delay time (Source delay) elapses.
• Triggered acquire (measurement) action
After the acquire trigger condition is satisfied, the channel performs current
and/or voltage measurement after the measurement delay time (Measure delay)
elapses.
Figure 6-9 Transient and Acquire Device Actions
Transient device action
- for voltage or current output -
Source delay
Acquire device action
- for voltage or current measurement -
Trigger
Step value
Step value
Measure Measure
Source delay
Measure delay Measure delay
Trigger
Trigger
Trigger