Agilent B2900 User’s Guide, Edition 2 5- 27
Front Panel Operations
Performing the Limit Test
11. In the /BUSY field, set the DIO pin number used for outputting the BUSY
signal.
12. In the /SOT field, set the DIO pin number used for outputting the start-of-test
(SOT) signal.
13. In the /EOT field, set the DIO pin number used for outputting the end-of-test
(EOT) signal.
14. To apply the settings, press Apply.
To apply the settings and close the dialog box, press OK.
To cancel the setting change, press the Cancel/Local key.
Setting Individual Limit Tests
For details on the setup parameters, see “Limit Test Setup dialog box” on page 4-32.
1. Press the Function > Limit Test > Limits function keys.
The Limit Test Setup dialog box is displayed.
2. In the Ch field, set the channel to which the setting change is applied.
3. In the Feed Data field, set the type of data used for the pass/fail judgement of the
limit test.
MATH: Calculation result data of math expression
VOLTS: Voltage measurement data
AMPS: Current measurement data
OHMS: Resistance data given by Resistance=Vmeas/Imeas
(Vmeas: Voltage measurement data, Imeas: Current measurement data)
4. In the Test Index field, set the index of the limit test between numbers 1 and 12.
Index numbers 1 to 12 are also used for bin numbers 1 to 12. For details, see
“Displaying the Limit Test Result” on page 5-29.
5. In the Limit Test field, set ON (enable) or OFF (disable) for the limit test
identified by Test Index.
6. In the Function field, set COMPLIANCE (compliance check) or LIMIT (limit
test) for the test mode.
7. When the operation mode is SORTING, set a bit pattern that represents the limit
test pass state (Pass) in the Pass Pattern field.