Agilent B2900 User’s Guide, Edition 2 5- 15
Front Panel Operations
Controlling the Source Output
The Output Filter dialog box is displayed.
2. Set each parameter. For details on the parameters, see “Output Filter dialog box”
on page 4-26.
3. To apply the settings, press Apply.
To apply the settings and close the dialog box, press OK.
To cancel the setting change, press the Cancel/Local key.
Setting the Connection Type
For details on this function, see “2-Wire Connections or 4-Wire Connections” on
page 3-11.
1. Press the Config > Source > Connection function keys.
The Output Connection dialog box is displayed.
2. In the Ch field, set the channel to which the setting change is applied.
3. Set 2-WIRE or 4-WIRE in the Sensing Type field.
4. To apply the settings, press Apply.
To apply the settings and close the dialog box, press OK.
To cancel the setting change, press the Cancel/Local key.
Setting the Low Terminal State
For details on this function, see “Floating” on page 3-12.
1. Press the Config > Source > Connection function keys.
The Output Connection dialog box is displayed.
2. In the Ch field, set the channel to which the setting change is applied.
3. Set GROUNDED or FLOATING in the Low Terminal State field.
4. To apply the settings, press Apply.
To apply the settings and close the dialog box, press OK.
To cancel the setting change, press the Cancel/Local key.