User`s guide
5- 6 Agilent B2900 User’s Guide, Edition 2
Front Panel Operations
Configuring Various Settings
Setting the Beeper
1. Press the More > System > More > Sound function keys.
2. To enable the beeper, press ON. To disable it, press OFF. To cancel the setting
change, press the Cancel/Local key.
Setting the Date and Time
1. Press the More > System > More > Info. function keys.
2. Press Date/Time.
The Date and Time dialog box is displayed.
3. Set the date and time.
4. To apply the settings, press Apply.
To apply the settings and close the dialog box, press OK.
To cancel the setting change, press the Cancel/Local key.
Performing the Self-Test
1. Press the On/Off switch and confirm that the switch is turned off.
2. Disconnect test leads and cables from the channel terminals.
3. Press the More > System > Cal/Test > Self-Test function keys.
The Confirmation dialog box is displayed.
4. Press OK to perform the self-test. To cancel the operation, press the
Cancel/Local key.
Performing the Self-Calibration
The self-calibration must be performed after warming-up of 60 minutes.
1. Press the On/Off switch and confirm that the switch is turned off.
2. Disconnect test leads and cables from the channel terminals.
3. Press the More > System > Cal/Test > Self-Cal function keys.
The Confirmation dialog box is displayed.