User`s guide

Agilent B2900 Users Guide, Edition 2 4- 37
Front Panel Reference
Trigger key group
Trigger Output Trigger output, ON or OFF
If this parameter is ON, B2900 sends an output trigger when it
changes the trigger state for the action specified by the Ch,
Layer, and Action parameters. See
Figure 6-8 on page 6-28.
Also see “DIO Configuration dialog box” on page 4-46 for
details on setting the trigger output timing.
The right field of the Trigger Output entry field shows the
present setting of the trigger output terminal. In the default
setting, it is the EXT1 terminal. It can be changed by the
following commands.
Between initial state and arm layer
:ARM[:ACQ|:TRAN]:TOUT:SIGN
Between arm layer and trigger layer
:TRIG[:ACQ|:TRAN]:TOUT:SIGN
Between trigger layer and transient action
:SOUR:TOUT:SIGN
Between trigger layer and acquire action
:SENS:TOUT:SIGN
Table 4-8 Trigger Source
Trigger Source Description
AUTO Signal internally generated and optimized for the present
operating mode
BUS Remote interface trigger command such as the group execute
trigger (GET) and the *TRG command
TIMER Signal internally generated every interval set by the Period
field
INT1 or INT2 Signal from the internal bus 1 or 2
LAN LXI trigger specified by the
:ARM[:ACQ|:TRAN]:SOUR:LAN and
:TRIG[:ACQ|:TRAN]:SOUR:LAN commands
EXTn Signal from the DIO pin n, which is an I/O port of the Digital
I/O D-sub connector on the rear panel. n=1 to 14