User`s guide
Agilent B2900 User’s Guide, Edition 2 4- 13
Front Panel Reference
Display and Assist Keys
Edit Opens the List Sweep dialog box used to set the list sweep
source.
Load Opens the Load List Sweep Data dialog box used to load a list
sweep data from a USB memory connected to the front panel
USB-A connector.
• List Sweep dialog box
This dialog box provides the following GUI for setting the list sweep source.
Data resolution is 6 digits on B2901A/B2902A, and 7 digits on
B2911A/B2912A.
(data graph) Displays the shape of the list sweep output
Type Data type V (voltage) or I (current)
CH Channel number 1 or 2, only on 2-channel models
(data list) Lists the data index and the output value
Points Number of data points
Max Maximum value
Min Minimum value
• Load List Sweep Data dialog box
This dialog box provides the following GUI for loading a list sweep data from a
file stored on a USB memory.
(data graph) Displays the shape of the list sweep data selected by the file
list
Path Folder into which the list sweep data file is saved
(file list) List of the list sweep data file
Points Number of data points
Max Maximum value
Min Minimum value
The following data can be loaded as the list sweep data.
• Comma separated value format, file extension csv
• Carriage return or line feed separated value format, file extension txt
• Space separated value format in a line, file extension prn