Technical data
3- 14 Agilent B2900A Configuration and Connection Guide, Edition 1
Connection
Connection with Prober
Table 3-1
Item number information for 4-wire (Kelvin) connection for the Low terminal
Kelvin to non-Kelvin connection
These instructions apply when the connections up to the Connector Plate are Kelvin, but
the probe is a non-Kelvin connection. Connect two Agilent 16494A triaxial cables between
the N1294A-002 banana to triaxial adapter for 4-wire and the connector plate. Connect the
triaxial connector on the probe cable as shown in Figure 3-23. Connect the FORCE and
SENSE lines on this side of the probe input terminal. A Tee Triaxial BNC adapter can be
used. For parts information, see Table 3-2.
NOTE To prevent oscillations, use triaxial cables as short as possible, if two 16494A triaxial cables with
Kelvin connections are used.
NOTE When connecting the cable and tee connector as shown in Figure 3-23, space restrictions make an
adjacent connector unusable. Since three connectors are necessary for a Kelvin connection a
connector plate must be used. Select the correct connector plate for your application.
Figure 3-23 Example of Kelvin to non-Kelvin connection
Table 3-2
Item number information for Kelvin to non-Kelvin connection with B2900A
Description Qty Product Number Part number
Triaxial Cable (1.5 m) - Max 1.0 A 1 16494A-001 N/A
Triaxial Cable (3.0 m) - Max 1.0 A 1 16494A-002 N/A
Triax (m) to Triax (f) adapter 1 N1254A-107 1250-2654
Description Qty Product Number Part number
Triaxial Cable (1.5 m) - Max 1.0 A 2 16494A-001 N/A
Triaxial Cable (3.0 m) - Max 1.0 A 2 16494A-002 N/A
Triaxial Cable (0.4 m) 1 16494A-004 N/A
Triaxial tee-connector 1 N/A 1250-1551
N1294A-002
Triaxial cable
16494A-001 1.5m
or
16494A-002 3.0m
High Sense
High Force
Low Force/
Low Sense
Connector plate
Shielding box
Probe cable (triaxial)
Tee triaxial BNC
1250-1551
Triaxial cable
16494A-004 0.4m
Prober
DUT