Technical data
Agilent B2900A Configuration and Connection Guide, Edition 1 3- 13
Connection
Connection with Prober
4-wire (Kelvin) connection for High Force and High Sense
These instructions apply when all connections are Kelvin. Two probes must contact the
wafer in this connection. Connect two Agilent 16494A triaxial cables between the High
Force and High Sense Terminals on the N1294A-002 banana to triaxial adapter for 4-wire
and the connector plate. Connect the FORCE and SENSE lines to probes separately.
Connect the triaxial connector on the probe cable as shown in Figure 3-21.
NOTE To prevent oscillations, use triaxial cables as short as possible, if two 16494A triaxial cables with
Kelvin connections are used.
Figure 3-21 Example of 4-wire (Kelvin) connection for High Force and High Sense
4-wire (Kelvin) connection for Low Force and Low Sense
Connect the triaxial connector using an adapter for GNDU (N1254A-107) as shown in
Figure 3-22. This adapter connects Force Terminal and Sense Terminal to the center
conductor. For parts information for this connection, see Table 3-1.
Figure 3-22 Example of 4-wire (Kelvin) connection for Low Force and Low Sense