Technical data
Agilent B2900A Configuration and Connection Guide, Edition 1 3- 5
Connection
Connection with Test Fixture
Figure 3-7 Agilent N1295A connection example for the BJT (NPN) by 2 channels of SMU
CAUTION Maximum voltage and current
The test fixture and adapters must be used under the following limitations to prevent damage to them.
Agilent N1294A-001 Banana to Triax adapter:
±250 V maximum, ±42 V maximum for connecting N1295A Small test fixture.
Agilent N1295A Small Test Fixture: ±42 V, 1.05 A maximum
NOTE Shielding
The N1295A is quipped with the lid. To minimize the affect of ambient noise, close the lid when
performing the measurement.
NOTE 4-wire connections are not supported on the N1295A and B2900A.
Low Force ch 1
High Force ch 1
Low Force ch 2
High Force ch 2
Emitter
Collector Base