User`s guide
1-12 Agilent B2200 User’s Guide, Edition 4
Introduction
Specifications
Switch Modules
Table 1-1 lists the specifications when the switch module is installed in the
mainframe. In the table, the conditions are as follows:
Conditions: 23 °C ± 5 °C, 5 % to 60 % R.H.
Table 1-1 Agilent B2210A/B2211A Switch Module Specifications
Agilent B2200A mainframe supports Agilent B2210A module.
Agilent B2201A mainframe supports Agilent B2211A module.
Mixed configuration of the switch modules is not supported.
B2210A B2211A
Maximum current rating (A)
I-V port 1.0 1.0
AUX port 0.5 0.5
Maximum voltage rating (V)
I-V port, channel to common 200 200
I-V port, channel to other channel 300 300
AUX port, channel to common 100 100
AUX port, channel to other channel 100 100
Closed channel residual resistance (Ω)
I-V port 0.6 0.6
AUX port 1.5 1.5
Channel isolation resistance (Ω)
I-V port
10
14
5 × 10
13
AUX port
10
9
10
9