User`s guide
Agilent B2200 User’s Guide, Edition 4 5-23
SCPI Command Reference
:DIAGnostic:TEST:FRAMe:STATe?
Syntax :DIAGnostic:TEST:FRAMe[:EXECute]? CONTroller | FPANel |
LED | PEN | BEEPer
Query response test_result <newline><^END>
1: fail
0: pass
Example OUTPUT @Agb2200;":DIAG:TEST:FRAM? CONT"
:DIAGnostic:TEST:FRAMe:STATe?
This command returns the most recent test result (pass/fail) of the specified test.
Syntax :DIAGnostic:TEST:FRAMe:STATe? CONTroller | FPANel | LED
| PEN | BEEPer
Query response test_result <newline><^END>
1: fail
0: pass
-1: not tested
Parameter Explanation
CONTroller Controller test
FPANel Front panel interface test
LED LED matrix test
PEN Light pen test
BEEPer Beeper test
Parameter Explanation
CONTroller Controller test
FPANel Front panel interface test
LED LED matrix test
PEN Light pen test
BEEPer Beeper test