Technical data

Module 13
SPGU Control and Applications
13-28
Flash Memory – Endurance Test
Measure Vth Connection
SMU3
Mech
Mech
Gate
Drain
Source
SMU1
Substrate
SMU4
PG4
Solid
State
Mech
Mech
SMU2
PG2
Solid
State
Mech
Mech
PG3
Selector or ASU ch2
Selector or ASU ch3
Selector or ASU ch1
Vth must be measured after writing or erasing. To perform the Vth measurement, all of the
selector/ASU channels must make the path to the SMU. Vth is measured at each decade, so a
mechanical relay will be used, not the solid state relay. The mechanical relay has low leakage