Technical data
Agilent 41000 Administration Guide, Edition 3 3- 31
Using Agilent iPACE Verification Tool
Reading Result Data
Table 3-4 Parameters in Report File
Parameter Description
Model Model number of the instrument that the self-test was performed.
Result Self-test result. Pass or Fail.
Source Source monitor unit or voltage source.
SMU1 to SMU8, VSU1 to VSU2, PGU1 to PGU2, or CMU.
Input
SWM input port number
1
that was used for the Source connection. 1
to 14.
1. SWM means the switching matrix.
Slot SWM slot number that was used for the Source connection. 1 to 4.
Output SWM output port number that was used for the Source connection. 1
to 12.
Judge Label of the button clicked by the operator after the connection check
was performed manually. Ok or Cancel.
Unit Voltmeter. SMU1 to SMU8, VMU1 to VMU2, or CMU.
MInput SWM input port number that was used for the Unit connection. 1 to
14.
MSlot SWM slot number that was used for the Unit connection. 1 to 4.
MOutput SWM output port number that was used for the Unit connection. 1 to
12.
VMeasure Voltage measurement data (V). Or the resistance measurement data
(Ω) if Unit is CMU.
Status Status data of Unit. 0: normal. For the status data, see manual of the
instrument (Unit).
Time Accumulated time from the beginning of the measurement (in
second).
IMeasure Current measurement data.
IOffset Averaging value of all measured current.
Noise Standard deviation value of all measured current.
ILeak Averaging value of the last ten measured current.
VStep Voltage output value from Source.