Technical data
Agilent 41000 Administration Guide, Edition 3 3- 25
Using Agilent iPACE Verification Tool
Using Batch Mode
Setup Example • Connection check:
All parameters must be specified. The program ignores the line that contains the
improper value.
Example: 0,SMU1,1,2,12,SMU4,5,2,9
Example: 0,SMU1,1,2,12,SMU4,5,2,7
Example: 0,SMU1,1,2,12,SMU4,5,2,5
Example: 0,SMU1,1,2,12,SMU4,5,2,3
Example: 0,SMU1,1,2,12,SMU4,5,2,1
• Noise offset measurement:
Specify the parameters test item to s output ch. The program ignores the line that
contains the improper value.
For the measurement without switching matrix, do not specify the parameters
s
input port, s output slot, and s output ch. Then you cannot delete comma.
Example: 1,SMU1,1,1,1
Example: 1,SMU2,2,1,2
Example: 1,SMU1,,,
• Settling time/leakage current measurement:
Specify the parameters test item to s output ch. The program ignores the line that
contains the improper value.
For the measurement without switching matrix, do not specify the parameters
s
input port, s output slot, and s output ch. Then you cannot delete comma.
Example: 2,SMU1,1,1,1
Example: 2,SMU2,2,1,2
Example: 2,SMU1,,,
• Self-test:
Specify test item only.
Example: 255
NOTE Sample Batch File
The following sample batch file is installed in the following folder.
folder: <installed folder>/bin
file name: iPACE Verification Tool (Batch).txt
The contents of the sample batch file is shown in Figure 3-12.