Technical data
3- 24 Agilent 41000 Administration Guide, Edition 3
Using Agilent iPACE Verification Tool
Using Batch Mode
To Create Batch File
This section explains how to create the batch file.
• “Batch File”
• “Connection Check”
• “To Create Batch File”
Especially, you have to understand the measurement mechanisms in the batch mode to
define the batch files to perform the connection check.
Batch File
Contents of the batch file are described in this section. Each line of the batch file should
contains the following information. The required parameter depends on the test item. The
parameters covered by [ ] can be ignored.
<test item>[,<s unit>,<s input port>,<s output slot>,<s output ch>[,
<unit>,<input port>,<output slot>,<output ch>]]
Table 3-3 Setup Parameters
Parameter
Description
1
1. Available values depend on the system component.
test item Test item. 0, 1, 2, or 255.
0: connection check
1: noise offset measurement
2: settling time/leakage current measurement
255: selftest
s unit Source monitor unit. SMU1 to SMU8.
For the connection check, s unit specifies the voltage source.
SMU1 to SMU8, VSU1 to VSU2, PGU1 to PGU2, or CMU.
s input port
SWM input port number
2
that is used for the s unit connection. 1 to 14.
2. SWM means the switching matrix.
s output slot SWM slot number that is used for the s unit connection. 1 to 4.
s output ch SWM output port number that is used for the s unit connection. 1 to 12.
unit Voltmeter. SMU1 to SMU8, VMU1 to VMU2, or CMU.
if CMU is set to s unit, set CMU to unit too. Then the CMU performs the
resistance measurement, not the voltage measurement.
input port SWM input port number that is used for the unit connection. 1 to 14.
output slot SWM slot number that is used for the unit connection. 1 to 4.
output ch SWM output port number that is used for the unit connection. 1 to 12.