Technical data
3- 22 Agilent 41000 Administration Guide, Edition 3
Using Agilent iPACE Verification Tool
Using Batch Mode
Using Batch Mode
This section explains how to use the batch mode that is the function to perform the
self-test, connection check, noise offset measurement, and settling time/leakage current
measurement automatically. To use the batch mode, you need a batch file that is used to
specify the measurement items and the measurement conditions.
• “Setup”
• “To Perform Batch Mode Verification”
• “To Create Batch File”
Setup
Make the following measurement setup to perform the verification by using the batch
mode. The following setup is not required for the self-test.
1. To perform the connection check:
Attach the E3143-60001 Connection Check Fixture to the Agilent E3140A Test Fixture
Adapter. See
Figure 3-10.
2. To perform the noise offset measurement or the settling/leakage current measurement:
• If you perform the measurement at the tip of the Agilent B2220A output pins:
Attach the E3190-60042 PV Open Fixture to the Agilent E3140A Test Fixture
Adapter. See
Figure 3-10.
• If you perform the measurement at the tip of the probe card:
Attach the probe card to the E3144-60001 card fixture. Then attach the fixture with
the probe card to the Agilent E3140A Test Fixture Adapter. See
Figure 3-10.
3. Mount the test fixture on the Agilent B2220A.
4. Connect the interlock cable between the DC source/monitor and the Agilent B2220A.
5. Close the lid of the test fixture adapter.
Figure 3-10 Measurement Setup