Technical data

3- 22 Agilent 41000 Administration Guide, Edition 3
Using Agilent iPACE Verification Tool
Using Batch Mode
Using Batch Mode
This section explains how to use the batch mode that is the function to perform the
self-test, connection check, noise offset measurement, and settling time/leakage current
measurement automatically. To use the batch mode, you need a batch file that is used to
specify the measurement items and the measurement conditions.
“Setup”
“To Perform Batch Mode Verification”
“To Create Batch File”
Setup
Make the following measurement setup to perform the verification by using the batch
mode. The following setup is not required for the self-test.
1. To perform the connection check:
Attach the E3143-60001 Connection Check Fixture to the Agilent E3140A Test Fixture
Adapter. See
Figure 3-10.
2. To perform the noise offset measurement or the settling/leakage current measurement:
If you perform the measurement at the tip of the Agilent B2220A output pins:
Attach the E3190-60042 PV Open Fixture to the Agilent E3140A Test Fixture
Adapter. See
Figure 3-10.
If you perform the measurement at the tip of the probe card:
Attach the probe card to the E3144-60001 card fixture. Then attach the fixture with
the probe card to the Agilent E3140A Test Fixture Adapter. See
Figure 3-10.
3. Mount the test fixture on the Agilent B2220A.
4. Connect the interlock cable between the DC source/monitor and the Agilent B2220A.
5. Close the lid of the test fixture adapter.
Figure 3-10 Measurement Setup