Technical data
Agilent 41000 Administration Guide, Edition 3 3- 19
Using Agilent iPACE Verification Tool
Settling Time/Leakage Current Measurement
Settling Time/Leakage Current Measurement
This section explains how to perform the settling time/leakage current measurement. The
settling time/leakage current measurement will be performed by applying voltage (10 V)
from the DC source/monitor and monitoring current by the DC source/monitor when the
Agilent B2220A outputs are open.
• “Measurement Setup”
• “Execution Procedure”
• “Troubleshooting”
Measurement Setup
Make the following measurement setup to perform the settling time/leakage current
measurement.
1. If you perform the measurement at the tip of the Agilent B2220A output pins:
Attach the E3190-60042 PV Open Fixture to the Agilent E3140A Test Fixture Adapter.
See
Figure 3-8.
2. If you perform the measurement at the tip of the probe card:
Attach the probe card to the E3144-60001 card fixture. Then attach the fixture with the
probe card to the Agilent E3140A Test Fixture Adapter. See
Figure 3-8.
3. Mount the test fixture on the Agilent B2220A.
4. Connect the interlock cable between the DC source/monitor and the Agilent B2220A.
5. Close the lid of the test fixture adapter.
Figure 3-8 Settling Time/Leakage Current Measurement Setup