Technical data

3- 16 Agilent 41000 Administration Guide, Edition 3
Using Agilent iPACE Verification Tool
Noise Offset Measurement
Noise Offset Measurement
This section explains how to perform the noise offset measurement. The noise offset
measurement will be performed by applying 0 V from the DC source/monitor and
monitoring current by the DC source/monitor when the Agilent B2220A outputs are open.
“Measurement Setup”
“Execution Procedure”
“Troubleshooting”
Measurement Setup
Make the following measurement setup to perform the noise offset measurement.
1. If you perform the measurement at the tip of the Agilent B2220A output pins:
Attach the E3190-60042 PV Open Fixture to the Agilent E3140A Test Fixture Adapter.
See
Figure 3-6.
2. If you perform the measurement at the tip of the probe card:
Attach the probe card to the E3144-60001 card fixture. Then attach the fixture with the
probe card to the Agilent E3140A Test Fixture Adapter. See
Figure 3-6.
3. Mount the test fixture on the Agilent B2220A.
4. Connect the interlock cable between the DC source/monitor and the Agilent B2220A.
5. Close the lid of the test fixture adapter.
Figure 3-6 Noise Offset Measurement Setup
3UREHU
$JLOHQW%$
$JLOHQW($
(RU
(
$JLOHQW($
3UREH&DUG,QWHUIDFH