Technical data
3- 4 Agilent 41000 Administration Guide, Edition 3
Using Agilent iPACE Verification Tool
Overview
Overview
The Agilent iPACE Verification Tool can perform the following test and measurements.
This tool will be helpful to specify the defective components of the Agilent 41000 or to
know the measurement ability of whole system including the extended measurement paths
and so on. It is also useful for the system administration.
• Self-test
Performs self-test of each instrument. You can see the test results on the Verification
Tool window or the log file.
• Connection check
Controls the specified instrument to force voltage. You can see the results on the
Verification Tool window or the log file.
This function will be helpful to specify the defective components by monitoring the
voltage at the end of the probe card interface, measurement cables, instrument
terminals, and so on.
• Noise offset measurement
Performs noise offset measurements. You can see the measurement results on the
Verification Tool window or the log file.
• Settling time/leakage current measurement
Performs settling time and leakage current measurements. You can see the
measurement results on the Verification Tool window or the log file.
• Automatic Execution
Performs any of the self-test, connection check, noise offset measurement,
settling
time/leakage current measurement automatically.
• Result Data Record
The result data will be logged into the log file. The default file is as follows.
folder: <installed folder>/bin
file name: iPACE Verification Tool (Report).txt
For example, <installed folder> will be as follows.
C:\Program Files\Agilent\iPACE Verification Tool A.01.00
NOTE To Change the Log File
To change the log file, use the Open Report File dialog box opened by clicking
File
> Open Report File. After changing the file name, the result data will be logged into
the new file.