Technical data

Agilent 41000 Administration Guide, Edition 3 2- 27
Installation and Operation
Maintenance
Maintenance
Maintenance should be performed periodically to keep the Agilent 41000 in good
condition.
Performance Verification
Performance verification must be performed periodically so that the instruments satisfy the
specifications, and keep a good condition. It is recommended to perform the performance
verification once a year at least. For the performance verification, contact your nearest
Agilent Technologies Service Center. Trained service personnel will perform the
performance verification.
Cleaning
Before performing cleaning, turn off each instrument of the Agilent 41000, and disconnect
power cable from each instrument’s rear panel. Use a dry cloth to clean the external case
parts.
To prevent electrical shock, do not perform cleaning when each instrument is turned on,
and do not use a wet cloth.
Model 300/400: Cleaning Contact Pins
To clean contact pins and probe card contacts, prepare the isopropyl alcohol and a dryer.
Then clean them as shown below.
1. Spray the isopropyl alcohol to pins or contacts.
2. Dry them enough and carefully using the dryer.
Do not wipe with a cloth. It may expand dirty area.