Technical data

2- 24 Agilent 41000 Administration Guide, Edition 3
Installation and Operation
Turning the PDU On and Off
Turning the PDU On and Off
This section describes how to turn on or off your Agilent 41000 equipped with the power
distribution unit (PDU).
“To Turn On your System”
“To Turn Off your System”
“To Turn Off your System in Emergency”
“To Reset EMO Button”
WARNING Electrical Work Type 3:
24 V is forced to the uninsulated parts of the EMO rear panel. Do not touch these
parts.
Figure 2-14 PDU and EMO
To Turn On your System
1. Set the line ON/OFF switch of the all instruments in the system cabinet to the ON
position.
2. Set the PDU instruments breaker to the ON position. The PDU is located on bottom of
the system cabinet.
3. Set the PDU controller breaker to the ON position.
4. Set the PDU main switch to the ON position.
5. Press the PDU system switch. Then the inside LED will turn on, which means AC
power is applied to the controller outlets that are the inside of the system cabinet.
6. Press the INSTRUMENT POWER ON button on the EMO panel. The green LED in
the button will turn on, which means AC power is applied to the instruments outlets
that are the inside of the system cabinet.