Technical data
Agilent 41000 Administration Guide, Edition 3 2- 7
Installation and Operation
Installation
Table 2-2 Model 100 Measurement Cable Connections
Table 2-3 Model 200 Measurement Cable Connections
Terminal name
Designation
on plate
1
1. Maximum connector number labelled depends on the connector plate.
Cable
Interlock Intlk Agilent 16493J Interlock Cable
GNDU GNDU Agilent 16493L GNDU Cable
E5270B
SMU
Force Odd number
from 1 to 24
For Kelvin connection. Agilent 16493K Kelvin Triaxial Cable.
16495F/G/H/J allow 6/12/3/4 Kelvin triaxial connections respectively.
Sense Even number
from 1 to 24
E5270B
SMU
Force 1 to 24 For non-Kelvin connection. Agilent 16494A Triaxial Cable.
16495F/G/H/J allow 12/24/6/8 triaxial connections respectively.
Sense - For non-Kelvin connection. Open Sense terminal.
4284A Unknown
(H
CUR
,
H
POT
,
L
CUR
,
L
POT
)
- Connect the Agilent 16048D/E Test Leads and connect it to the
Agilent E5288A Atto Sense/Switch Unit (ASU). See
“Model 100: To
Connect ASU” on page 2-4. Two each of the ASUs are required.
Or connect the test leads and connect it to the AUX terminals (coaxial
BNC) of the 16495H/16495J connector plate. Then the Agilent
16494F CMU Cable can be used.
Terminal name
Designation
on plate
1
1. Maximum connector number labelled depends on the connector plate.
Cable
Interlock Intlk Agilent 16493J Interlock Cable
GNDU GNDU Agilent 16493L GNDU Cable
B2200A
Output
Force Odd number
from 1 to 24
For Kelvin connection. Agilent 16494B Kelvin Triaxial Cable.
16495F/G allow 6/12 Kelvin triaxial connections respectively.
Sense Even number
from 1 to 24
B2200A
Output
Force 1 to 24 For non-Kelvin connection. Agilent 16494A Triaxial Cable.
16495F/G allow 12/24 triaxial connections respectively.
Sense - For non-Kelvin connection. Open Sense terminal.
4284A Unknown
(H
CUR
,
H
POT
,
L
CUR
,
L
POT
)
- Connect the Agilent 16494F CMU cable and connect it to the Agilent
B2200A Input 13 (CMH) and 14 (CML).