Technical data

2- 4 Agilent 41000 Administration Guide, Edition 3
Installation and Operation
Installation
Model 100: To Connect ASU
The Agilent E5288A Atto Sense and Switch Unit (ASU) will add the 1 pA range to the
high resolution SMU (HRSMU) when it is connected between the HRSMU and the device
under test (DUT). And the ASU provides the input selection function when instruments are
connected to the SMU input and the AUX input.
NOTE For the installation of the ASU and the connection to the DUT interface, contact your
favorite prober vender. The prober vender will be able to provide the DUT interface that
can be connected directly to the ASU output. Dimensions of the ASU are 132 mm (W)
×
50 mm (D) × 88.5 mm (H) excluding the connectors.
Figure 2-1 Image of ASU Installation
To install 16495K Fix the Agilent 16495K Plate to the shielding box or something that will cover the
manipulators or the probe card (DUT interface). See Agilent 16495 Installation Guide. The
16495K is the plate used to block the light from the cable hole.
To install ASU 1. Fix the ASU in the shielding box. The ASU must be fixed to the best position for
accessing cables that come from the DUT interface and the instruments. The Agilent
N1254A-108 Magnetic Stand will be useful for fixing the ASU.
2. Pass the measurement cables (Agilent 16493M, 16494A, BNC cable, and so on)
through the cable hole of the 16495K. They will be connected between the ASU and
the instruments
3. Connect the 16493M cable to the ASU D-sub connector.
Connect the 16494A triaxial cable to the ASU Force terminal.
Connect the BNC cable to the ASU AUX In terminal. For the instrument without the
four-terminal-pairs inputs, open the CMU-pot terminal or cover it by using the BNC
open cap.
4. Adjust the cable length in the shielding box, and catch in the cables by using the
16495K.
5. Connect the cable extended from the DUT interface to the ASU output terminal. If you
do not make the Kelvin connection, open the Sense terminal or cover it by using the
triaxial open cap.
Prober
Shielding box
ASU
ASU
DUT interface
To E5270B/4284A
16495K
16495K
To E5270B/4284A