Technical data

Agilent 41000 Administration Guide, Edition 3 1- 9
Introduction
System Cabinet
System Cabinet
This section explains important panel features of the Agilent 41000 system cabinet. See
Table 1-6 for the EMO/PDU operation and response.
Figure 1-2 Emergency Power Off (EMO) Panel
EMO Button
EMO button is used to turn off the power distribution unit (PDU) immediately. If the EMO
button is normal position (out position), AC power can be applied to the PDU. But if the
EMO button is pressed (set to the in position), AC power is shut off and the LINE indicator
is turned off.
LINE Indicator
LINE indicator lights when AC power is applied to the PDU.
INSTRUMENT POWER ON Button
INSTRUMENT POWER ON button is used to apply AC power to the instruments outlets
that are the inside of the system cabinet. If you press this button when the LINE indicator is
ON, AC power will be applied to the instruments outlets and the inside green LED will be
turned on.
INSTRUMENT POWER OFF Button
INSTRUMENT POWER OFF button is used to shut off AC power to the instruments
outlets. If you press this button, AC power to the instruments outlets will be shut off.
'/1DWVVQP
+05647/'06219'4
1((10
.+0'KPFKECVQT