Measurement Cable Connections
Table 7-15 Agilent 4156C Connections with HPSMU
NOTE To make Sense terminal open
Connect the triaxial open connector (Agilent part number 1250-1708) to the Sense
terminal.
Instrument
terminal
1
1. For the non-Kelvin connection, connect the Force terminal and open the Sense
terminal.
Cable B2200 Input
4156C SMU1 Agilent 16494A triaxial cable 1
Agilent 16493K Kelvin triaxial cable
2
2. with 2 or more Kelvin options.
1 and 2
4156C SMU2 Agilent 16494A triaxial cable 2 or 3
Agilent 16493K Kelvin triaxial cable
3
3. with 3 or 4 Kelvin options.
3 and 4
4156C SMU3 Agilent 16494A triaxial cable 3 or 4 or 5
Agilent 16493K Kelvin triaxial cable
4
4. with 4 Kelvin options.
5 and 6
4156C SMU4 Agilent 16494A triaxial cable 4 or 5 or 6 or open
4156C VSU1
5
5. VSU is not connected if the PGU option is installed.
Agilent 16493B coaxial cable 9
4156C VSU2
5
Agilent 16493B coaxial cable 10
4156C VMU1 Agilent 16493B coaxial cable 11
4156C VMU2 Agilent 16493B coaxial cable 12
4156C Interlock Agilent 16493J Interlock cable to B2220A interlock
41501B HPSMU Agilent 16494A triaxial cable 5
Agilent 16493K Kelvin triaxial cable
6
6. with 1 or more Kelvin options.
7 and 8
41501B PGU1 Agilent 16493B coaxial cable 9
41501B PGU2 Agilent 16493B coaxial cable 10