Service
Replacement Procedure
To Remove EMO Panel
1. Set the main switch on the power distribution unit (PDU) to the OFF position.
2. Disconnect the PDU system power cable from the power outlet of the installation site.
3. Remove the top panel. Then see “To Remove Top Panel” on page 7-21.
4. Remove the top cover. Then see “To Remove Top Cover” on page 7-22.
5. Disconnect the EMO cable from the rear side of the PDU.
6. Remove the nut that fastens the EMO ground wire to the system cabinet. Then use a nut
driver (H8).
7. Remove screws at the bottom of the EMO panel. Then use a plus screwdriver (H2).
8. Remove the EMO panel.
Figure 7-14 Removing EMO Panel