Technical data

Agilent 41000 Administration Guide, Edition 3 5- 19
Measurement Techniques
Capacitance Measurements
Capacitance Measurements
NOTE Information described in this section is effective for the Agilent 41000 Model 200/300/400.
When the capacitance/conductance measurement is performed, LCR meter measures the
capacitance/conductance of the path including a device under test (DUT), matrix switches,
extension cables and so on. So, the data measured by the LCR meter is far from the DUT’s
capacitance/conductance.
The Agilent B2200 VXIplug&play driver provides the functions used to compensate the
capacitance/conductance measured by the Agilent 4284A LCR meter in the measurement
environments described in
“Required Conditions” on page 5-20.
This section explains how to use the capacitance compensation function.
“Capacitance Compensation Function”
“Required Conditions”
“To Create Compensation Data File”
“To Perform Measurement and Compensation”
“agb220xa_compenC”
“agb220xa_selectCompenFile”
Capacitance Compensation Function
Driver functions used for the capacitance compensation are listed below.
agb220xa_selectCompenFile function
agb220xa_compenC function
NOTE Corrected data by the function is not guaranteed. But typical data (supplemental data) is as
follows.
Capacitance measurement accuracy (typical): ±1 % ±0.5 pF
This typical data is for the following measurement conditions:
Measurement frequency: 1 kHz to 1 MHz
Measurement range: Maximum 1000 pF
Measurement terminal: At the end of the Agilent 16494A/B/C cable connected to the
switch module output terminals.
The typical data does not apply to anything extended from the 16494A/B/C cable. The
conditions described in “Required Conditions” on page 5-20 must be satisfied.