Technical data

Agilent 41000 Administration Guide, Edition 3 5- 17
Measurement Techniques
Low Resistance Measurements
To Select the Output Current Value
The current source output value is a key factor to perform resistance measurement
accurately. The optimum value can be find as shown below.
Summary Performs the voltage/current measurement for a sample with current output log sweep
Plots the resistance, measured voltage, measured current vs. output current curves
Plots the resistance vs. measured voltage plot curve
Plots the resistance vs. measured current plot curve
Decides the current output value for the spot resistance measurement
Procedure 1. Connect a sample of the device under test (DUT). And perform current output log
sweep (e.g. Iforce=1 µA to 100 mA). Then measure voltage (Vmeas) and current
(Imeas).
2. Calculate resistance (Rcalc). And plot the Vmeas, Imeas, Rcalc vs. Iforce
characteristics on the log-log graph as shown in
Figure 5-10.
3. Plot the Rcalc vs. Imeas, Vmeas characteristics on the linear-log graph as shown in
Figure 5-11.
4. Decide the current source output value for the resistance spot measurement.
For example of Figure 5-11, the resistance value is stable in the range over 2 mA. The
variation is about 1 % of the resistance value (about 35.55 m). So, an output current
value should be selected from this range. And the other one should be about 1/10 of the
selected current, or less.
Figure 5-10 Vmeas, Imeas, Rcalc vs. Iforce Plot Example
1.E-06
1.E-05
1.E-04
1.E-03
1.E-02
1.E-01
1.E-06 1.E-05 1.E-04 1.E-03 1.E-02
Iforce
Vmeas or Imeas or Rcalc
Vmeas
Imeas
Rcalc
Rcalc =
Vm Vm(at If=1 µA)
Im Im(at If=1 µA)