Technical data
4- 12 Agilent 41000 Administration Guide, Edition 3
Agilent iPACE Verification Tool
To Change Measurement Conditions
Table 4-5 Measurement Condition Setup Parameters
Var i abl e n am e Description Default value
Connection check
VERIFY_CHECK_VS_FORCE
Voltage source output value (V) 10.0
VERIFY_CHECK_VS_COMPLIANCE
Voltage source compliance value (A) 0.001
VERIFY_CHECK_IS_FORCE
Voltage measurement unit output value (A) 0.0
VERIFY_CHECK_IS_COMPLIANCE
Voltage measurement unit compliance value (V) 20.0
Noise offset measurement
VERIFY_NOISE_INTEG_VALUE
Integration time (number of power line cycles) 50
VERIFY_NOISE_VS_FORCE
Voltage source output value (V) 0.0
VERIFY_NOISE_VS_COMPLIANCE
Voltage source compliance value (A) 0.001
VERIFY_NOISE_VS_NUMBER
Number of measurement points 50
VERIFY_NOISE_HOLD_TIME
Hold time (1.0 = 1 s) 10.0
Settling time/leakage current measurement
VERIFY_SETTLE_INTEG_VALUE
Integration time (number of power line cycles) 2
VERIFY_SETTLE_VS_FORCE
Voltage source output value (V) 10.0
VERIFY_SETTLE_VS_COMPLIANCE
Voltage source compliance value (A) 0.001
VERIFY_SETTLE_IM_TIME
Total measurement time (repetition time, 10000.0 = 1 s) 1000000.0
VERIFY_SETTLE_HOLD_TIME
Hold time (1.0 = 1 s) 0.0
VERIFY_LEAK_VS_NUMBER
Number of measurement data used to calculate the
ILeak value
10