Technical data

Agilent 41000 Administration Guide, Edition 3 4- 11
Agilent iPACE Verification Tool
To Change Measurement Conditions
To Change Measurement Conditions
The following procedure provides how to change the measurement condition and the
instrument setup. See
Table 4-5 for the measurement condition setup parameters.
1. Make backup of all files in <installed folder>\src.
For example, copy them to <installed folder>\backup\src.
Then, <installed folder> is the folder where this program has been installed.
For
example,
C:\Program Files\Agilent\iPACE Verification Tool A.01.00
2. Change the name of the following files. Add .org after .dll.
Folder: <installed folder>\bin
File: Agilent.IpaceVerificationControl.dll
File: Agilent.IpaceVerificationInstrument.dll
3. Open the following file on Microsoft Visual Basic .Net software.
Folder: <installed folder>\src\iPACE Verification Tool
File: Parameter.vb
4. Change the value of the parameters defined in the file as you want.
For the variables (measurement condition setup parameters), see Table 4-5.
5. Build the following projects in this order.
a. Instrument project
b. Control project
c. Body project
d. Tool project
6. Copy the following files to <installed folder>\bin.
Folder: <installed folder>\src\iPACE Verification Body\bin
File: Agilent.IpaceVerificationControl.dll
File: Agilent.IpaceVerificationInstrument.dll