User`s guide
Table Of Contents
- User’s Guide
- 1 Getting Started
- 2 Introduction
- 3 Installation
- 4 Using EasyEXPERT
- 5 Classic Test Definition
- I/V Sweep
- Multi Channel I/V Sweep
- I/V List Sweep
- I/V-t Sampling
- C-V Sweep
- Direct Control
- Function Setup
- Auto Analysis Setup
- Display Setup
- SMU Range Setup Window
- ADC and Integration Time Setup Window
- Advanced Setup Window
- CMU Range Setup Window
- Advanced Setup Window for C-V Sweep
- Switching Matrix Control
- SPGU Control
- SPGU Pulse Setup Window
- Load Z Setup Window
- Pulse Switch Setup Window
- SPGU ALWG Setup Window
- Define ALWG Waveform Window
- 6 Application Test Definition
- 7 Function Details
- I/V Sweep Measurement
- Multi Channel I/V Sweep Measurement
- I/V-t Sampling Measurement
- C-V Sweep Measurement
- SPGU Module
- Sweep Abort Function
- Standby Function
- Bias Hold Function
- Current Offset Cancel
- SMU CMU Unify Unit
- Atto Sense and Switch Unit
- SMU/PG Selector
- SMU Ranging Mode
- SMU Compliance
- SMU Pulse
- SMU Measurement Time
- SMU Filter
- SMU Series Resistor
- Interlock Function
- Auto Power Off Function
- Initial Settings
- 8 Built-in Programming Tool
- 9 If You Have a Problem
- When You Operate B1500A
- When You Perform Measurement
- Measurement Takes More Time than Specified
- Noise Affects the Measured Values
- Voltage Measurement Error is Large
- SMU Oscillates for High-Frequency Device Measurements
- SMU Oscillates for Negative Resistance Measurements
- Large Current Causes High Temperature (Thermal Drift)
- Measurement Damages the Device under Test
- Leaving Connections Damages Devices after Measurement
- Unexpected Sampling Measurement Data is Returned
- MFCMU Causes Unbalance Condition
- Before Shipping to Service Center
- Data Backup and Recovery
- B1500A System Recovery
- Updating EasyEXPERT
- Error Codes
- 10 Application Library and Utilities

2- 16 Agilent B1500 User’s Guide, Edition 7
Introduction
Measurement Units
Measurement Units
Agilent B1500A can be equipped with the following measurement modules.
• “GNDU - Ground Unit”
• “B1510A High Power SMU”
• “B1511A Medium Power SMU”
• “B1517A High Resolution SMU”
• “B1520A Multi Frequency CMU”
• “B1525A High Voltage SPGU”
• “B1530A WGFMU”
See “Using EasyEXPERT” on page 4-1 for the functions available for Agilent
EasyEXPERT software. See Programming
Guide for the functions available for the
GPIB remote mode. See Agilent B1530A User’s Guide for using the WGFMU.
GNDU - Ground Unit
Agilent B1500A is equipped with the ground unit (GNDU). The GNDU is a 0 V
constant voltage source, and used for the reference of measurement ground. Also the
GNDU can sink up to ± 4.2 A, so it is effective for using the HPSMU (high power
SMU). Figure 2-1 is a simplified circuit diagram of GNDU.
Figure 2-1 Simplified GNDU Circuit Diagram
*1'8
9
)RUFH
&RPPRQ