User`s guide
Table Of Contents
- User’s Guide
- 1 Getting Started
- 2 Introduction
- 3 Installation
- 4 Using EasyEXPERT
- 5 Classic Test Definition
- I/V Sweep
- Multi Channel I/V Sweep
- I/V List Sweep
- I/V-t Sampling
- C-V Sweep
- Direct Control
- Function Setup
- Auto Analysis Setup
- Display Setup
- SMU Range Setup Window
- ADC and Integration Time Setup Window
- Advanced Setup Window
- CMU Range Setup Window
- Advanced Setup Window for C-V Sweep
- Switching Matrix Control
- SPGU Control
- SPGU Pulse Setup Window
- Load Z Setup Window
- Pulse Switch Setup Window
- SPGU ALWG Setup Window
- Define ALWG Waveform Window
- 6 Application Test Definition
- 7 Function Details
- I/V Sweep Measurement
- Multi Channel I/V Sweep Measurement
- I/V-t Sampling Measurement
- C-V Sweep Measurement
- SPGU Module
- Sweep Abort Function
- Standby Function
- Bias Hold Function
- Current Offset Cancel
- SMU CMU Unify Unit
- Atto Sense and Switch Unit
- SMU/PG Selector
- SMU Ranging Mode
- SMU Compliance
- SMU Pulse
- SMU Measurement Time
- SMU Filter
- SMU Series Resistor
- Interlock Function
- Auto Power Off Function
- Initial Settings
- 8 Built-in Programming Tool
- 9 If You Have a Problem
- When You Operate B1500A
- When You Perform Measurement
- Measurement Takes More Time than Specified
- Noise Affects the Measured Values
- Voltage Measurement Error is Large
- SMU Oscillates for High-Frequency Device Measurements
- SMU Oscillates for Negative Resistance Measurements
- Large Current Causes High Temperature (Thermal Drift)
- Measurement Damages the Device under Test
- Leaving Connections Damages Devices after Measurement
- Unexpected Sampling Measurement Data is Returned
- MFCMU Causes Unbalance Condition
- Before Shipping to Service Center
- Data Backup and Recovery
- B1500A System Recovery
- Updating EasyEXPERT
- Error Codes
- 10 Application Library and Utilities

Agilent B1500 User’s Guide, Edition 7 10-57
Application Library and Utilities
Utility Programs
XSLT
The XSLT (XSL-Transformation, eXtensible Stylesheet Language Transformation)
filter files are used to export test records. The fifteen filter files listed below are
stored in the C:\Program Files\Agilent\B1500\EasyEXPERT\Utilities\XSLT folder.
See Table 10-7.
Filter files that contain csv in the file name export test record in the format that uses
a comma (,) as data separator.
Filter files that contain tab-sv in the file name export test record in the format that
uses a tab as data separator.
Filter files that contain xmlss in the file name export test record in the XML Spread
Sheet format.
To export test
record
Test record can be exported by the following procedure.
1. Open the Export in My Format (p. 4-22) dialog box by clicking Results >
Transport Data > Export in My Format....
2. Enter the XSLT filter file name, and click the Export As... button.
3. On the Export As dialog box, specify the file name and file type for the export
data, and click Save.
To read export
result
The exported data file can be opened and read with a spreadsheet software. The
procedure for opening the export data with a spreadsheet software is as follows.
• For data separated by a comma
Specify the data separated by a comma for the original data format.
• For data separated by a tab
Specify the data separated by a tab for the original data format.
• For data in the XML Spread Sheet format
Specify the XML file for the file type.
After reading the file, all data in the file will be entered into independent cells on the
spreadsheet software.