User`s guide
Table Of Contents
- User’s Guide
- 1 Getting Started
- 2 Introduction
- 3 Installation
- 4 Using EasyEXPERT
- 5 Classic Test Definition
- I/V Sweep
- Multi Channel I/V Sweep
- I/V List Sweep
- I/V-t Sampling
- C-V Sweep
- Direct Control
- Function Setup
- Auto Analysis Setup
- Display Setup
- SMU Range Setup Window
- ADC and Integration Time Setup Window
- Advanced Setup Window
- CMU Range Setup Window
- Advanced Setup Window for C-V Sweep
- Switching Matrix Control
- SPGU Control
- SPGU Pulse Setup Window
- Load Z Setup Window
- Pulse Switch Setup Window
- SPGU ALWG Setup Window
- Define ALWG Waveform Window
- 6 Application Test Definition
- 7 Function Details
- I/V Sweep Measurement
- Multi Channel I/V Sweep Measurement
- I/V-t Sampling Measurement
- C-V Sweep Measurement
- SPGU Module
- Sweep Abort Function
- Standby Function
- Bias Hold Function
- Current Offset Cancel
- SMU CMU Unify Unit
- Atto Sense and Switch Unit
- SMU/PG Selector
- SMU Ranging Mode
- SMU Compliance
- SMU Pulse
- SMU Measurement Time
- SMU Filter
- SMU Series Resistor
- Interlock Function
- Auto Power Off Function
- Initial Settings
- 8 Built-in Programming Tool
- 9 If You Have a Problem
- When You Operate B1500A
- When You Perform Measurement
- Measurement Takes More Time than Specified
- Noise Affects the Measured Values
- Voltage Measurement Error is Large
- SMU Oscillates for High-Frequency Device Measurements
- SMU Oscillates for Negative Resistance Measurements
- Large Current Causes High Temperature (Thermal Drift)
- Measurement Damages the Device under Test
- Leaving Connections Damages Devices after Measurement
- Unexpected Sampling Measurement Data is Returned
- MFCMU Causes Unbalance Condition
- Before Shipping to Service Center
- Data Backup and Recovery
- B1500A System Recovery
- Updating EasyEXPERT
- Error Codes
- 10 Application Library and Utilities

Agilent B1500 User’s Guide, Edition 7 9-73
If You Have a Problem
Error Codes
N807 CMU failed Vch full scale measurement.
N808 CMU failed nominal gain measurement
N809 CMU failed extent range X3 adjustment.
N810 CMU failed range resistor 50ohm adjustment.
N811 CMU failed range resistor 1kohm adjustment.
N812 CMU failed range resistor 10kohm adjustment.
N813 CMU failed range resistor 100kohm adjustment.
N814 CMU failed relative Z calculation.
N820 CMU failed correction.
N830 CMU failed configuration test.
N831 SCUU failed SCUU configuration test.
N832 SCUU failed SMU configuration test.
N833 SCUU failed CMU configuration test.
N834 CMU failed digital function test.
N835 CMU failed CPLD test.
N836 CMU failed FPGA test.
N837 CMU failed EEPROM test.
N838 CMU failed PLL1/PLL2 test.
N839 CMU failed PLL DET low state test.
N840 CMU failed PLL DET high state test.
N841 CMU failed PLL1 lock test
N842 CMU failed PLL2 lock test.
N843 CMU failed PLL2 lock test.
N844 CMU failed Hcur DC and VRD ADC test.
N845 CMU failed DC offset test.
N846 CMU failed DC bias 0V test.
N847 CMU failed DC bias -25V test.
N848 CMU failed DC bias +25V test.