User`s guide
Table Of Contents
- User’s Guide
- 1 Getting Started
- 2 Introduction
- 3 Installation
- 4 Using EasyEXPERT
- 5 Classic Test Definition
- I/V Sweep
- Multi Channel I/V Sweep
- I/V List Sweep
- I/V-t Sampling
- C-V Sweep
- Direct Control
- Function Setup
- Auto Analysis Setup
- Display Setup
- SMU Range Setup Window
- ADC and Integration Time Setup Window
- Advanced Setup Window
- CMU Range Setup Window
- Advanced Setup Window for C-V Sweep
- Switching Matrix Control
- SPGU Control
- SPGU Pulse Setup Window
- Load Z Setup Window
- Pulse Switch Setup Window
- SPGU ALWG Setup Window
- Define ALWG Waveform Window
- 6 Application Test Definition
- 7 Function Details
- I/V Sweep Measurement
- Multi Channel I/V Sweep Measurement
- I/V-t Sampling Measurement
- C-V Sweep Measurement
- SPGU Module
- Sweep Abort Function
- Standby Function
- Bias Hold Function
- Current Offset Cancel
- SMU CMU Unify Unit
- Atto Sense and Switch Unit
- SMU/PG Selector
- SMU Ranging Mode
- SMU Compliance
- SMU Pulse
- SMU Measurement Time
- SMU Filter
- SMU Series Resistor
- Interlock Function
- Auto Power Off Function
- Initial Settings
- 8 Built-in Programming Tool
- 9 If You Have a Problem
- When You Operate B1500A
- When You Perform Measurement
- Measurement Takes More Time than Specified
- Noise Affects the Measured Values
- Voltage Measurement Error is Large
- SMU Oscillates for High-Frequency Device Measurements
- SMU Oscillates for Negative Resistance Measurements
- Large Current Causes High Temperature (Thermal Drift)
- Measurement Damages the Device under Test
- Leaving Connections Damages Devices after Measurement
- Unexpected Sampling Measurement Data is Returned
- MFCMU Causes Unbalance Condition
- Before Shipping to Service Center
- Data Backup and Recovery
- B1500A System Recovery
- Updating EasyEXPERT
- Error Codes
- 10 Application Library and Utilities

9-64 Agilent B1500 User’s Guide, Edition 7
If You Have a Problem
Error Codes
3312 Specified ALWG measurement interval time is out of absolute limits.
Check the measurement interval time and set the correct value.
The
value must be 10 ns to 1.34217728 s, in 10 ns resolution.
3313 Specified ALWG measurement instruction code is invalid.
Check the measurement event setting and set the correct values.
3314 Specified ALWG range change instruction code is invalid.
Check the range event setting and
set the correct values
.
3315 Specified ALWG measurement count is out of ab
solute limits.
Check the measurement averaging tim
e and set the correct value. The
value must be 0, or 10 ns to 0.020971512 s, in 10 ns resolution.
3316 Specified ALWG measurement count is greater than measurement
in
terval.
Check the measurement averaging tim
e and set the correct value. The
value must less than or equal to the measurement interval time.
3317 Specified slot is invalid.
Check the slot number and set the correct value. The slot number must
be 1 to 10.
3318
Specified module channel is invalid.
Check the channel number and
set the correct value.
3319
Output delay is out of absolute limits.
Check the output delay and set the correct value. The value must be
-50 ns to
50 ns, in 625 ps resolution.
3320 Measurement delay is out of absolute limits.
Check the measurement delay and set the correct val
ue. The value must
be -50 ns to 50 ns, in 625 ps resolution.
3321 VM/IM measurement mode is invalid.
Check the measurement mode and set the corr
ect value.
3322 Voltage measurement range is invalid.
Check the voltage measurement range and set the correct value.
3323 Current measurement range is invalid.