User`s guide
Table Of Contents
- User’s Guide
- 1 Getting Started
- 2 Introduction
- 3 Installation
- 4 Using EasyEXPERT
- 5 Classic Test Definition
- I/V Sweep
- Multi Channel I/V Sweep
- I/V List Sweep
- I/V-t Sampling
- C-V Sweep
- Direct Control
- Function Setup
- Auto Analysis Setup
- Display Setup
- SMU Range Setup Window
- ADC and Integration Time Setup Window
- Advanced Setup Window
- CMU Range Setup Window
- Advanced Setup Window for C-V Sweep
- Switching Matrix Control
- SPGU Control
- SPGU Pulse Setup Window
- Load Z Setup Window
- Pulse Switch Setup Window
- SPGU ALWG Setup Window
- Define ALWG Waveform Window
- 6 Application Test Definition
- 7 Function Details
- I/V Sweep Measurement
- Multi Channel I/V Sweep Measurement
- I/V-t Sampling Measurement
- C-V Sweep Measurement
- SPGU Module
- Sweep Abort Function
- Standby Function
- Bias Hold Function
- Current Offset Cancel
- SMU CMU Unify Unit
- Atto Sense and Switch Unit
- SMU/PG Selector
- SMU Ranging Mode
- SMU Compliance
- SMU Pulse
- SMU Measurement Time
- SMU Filter
- SMU Series Resistor
- Interlock Function
- Auto Power Off Function
- Initial Settings
- 8 Built-in Programming Tool
- 9 If You Have a Problem
- When You Operate B1500A
- When You Perform Measurement
- Measurement Takes More Time than Specified
- Noise Affects the Measured Values
- Voltage Measurement Error is Large
- SMU Oscillates for High-Frequency Device Measurements
- SMU Oscillates for Negative Resistance Measurements
- Large Current Causes High Temperature (Thermal Drift)
- Measurement Damages the Device under Test
- Leaving Connections Damages Devices after Measurement
- Unexpected Sampling Measurement Data is Returned
- MFCMU Causes Unbalance Condition
- Before Shipping to Service Center
- Data Backup and Recovery
- B1500A System Recovery
- Updating EasyEXPERT
- Error Codes
- 10 Application Library and Utilities

9-60 Agilent B1500 User’s Guide, Edition 7
If You Have a Problem
Error Codes
Set the appropriate count value to SPRM.
2105 Specified load voltage is out of range.
Set the appropriate voltage to SPV or ALW.
2106 Specified load voltage of added amplitude is out of range.
Set the appropriate voltage to SPV for setting the 3-level pulse output.
2107 Specified voltage is out of absolute limits (can't achieve amplitude).
Set the appropriate voltage to SPV. It must be the voltage which can be
ap
plied under the presen
t load impedance condition.
2108 Specified transition time is out of absolute limits.
Set the appropriate value of leading time or
trailing time to SPT.
2111 Leading/0.8 must be within Width value.
Set the appropriate value to SPT. The pulse width value must be more
than leadin
g time/0.8 value.
2112 Delay + Width + Trail/0.8 must be within Period value.
Set the appropriate value to SPT. The pulse period value must be more
than delay ti
me+pulse width+trailing time/0.8 value.
2113 Specified pulse delay is out of absolute limits.
Set the appropriate delay time value to SPT.
2114 Specified pulse width is out of absolute limits.
Set the appropriate pulse width value to SPT.
2115 Specified duration is out of absolute limits.
Set the appropriate duration value to SPRM.
2121 Delay + Width must be within Period value (ODSW Timing).
Set the appropriate value to ODSW. The period value must be more
th
an delay+widt
h value.
2122 Specified ODSW delay timing parameter out of absolute limits.
Set the appropriate delay value to
ODSW.
2123 Specified ODSW width timing parameter out of absolute limits.
Set the appropriate width value to
ODSW.
2131 Delay + Interval * N must be within Period value (ADC Timing).