User`s guide
Table Of Contents
- User’s Guide
- 1 Getting Started
- 2 Introduction
- 3 Installation
- 4 Using EasyEXPERT
- 5 Classic Test Definition
- I/V Sweep
- Multi Channel I/V Sweep
- I/V List Sweep
- I/V-t Sampling
- C-V Sweep
- Direct Control
- Function Setup
- Auto Analysis Setup
- Display Setup
- SMU Range Setup Window
- ADC and Integration Time Setup Window
- Advanced Setup Window
- CMU Range Setup Window
- Advanced Setup Window for C-V Sweep
- Switching Matrix Control
- SPGU Control
- SPGU Pulse Setup Window
- Load Z Setup Window
- Pulse Switch Setup Window
- SPGU ALWG Setup Window
- Define ALWG Waveform Window
- 6 Application Test Definition
- 7 Function Details
- I/V Sweep Measurement
- Multi Channel I/V Sweep Measurement
- I/V-t Sampling Measurement
- C-V Sweep Measurement
- SPGU Module
- Sweep Abort Function
- Standby Function
- Bias Hold Function
- Current Offset Cancel
- SMU CMU Unify Unit
- Atto Sense and Switch Unit
- SMU/PG Selector
- SMU Ranging Mode
- SMU Compliance
- SMU Pulse
- SMU Measurement Time
- SMU Filter
- SMU Series Resistor
- Interlock Function
- Auto Power Off Function
- Initial Settings
- 8 Built-in Programming Tool
- 9 If You Have a Problem
- When You Operate B1500A
- When You Perform Measurement
- Measurement Takes More Time than Specified
- Noise Affects the Measured Values
- Voltage Measurement Error is Large
- SMU Oscillates for High-Frequency Device Measurements
- SMU Oscillates for Negative Resistance Measurements
- Large Current Causes High Temperature (Thermal Drift)
- Measurement Damages the Device under Test
- Leaving Connections Damages Devices after Measurement
- Unexpected Sampling Measurement Data is Returned
- MFCMU Causes Unbalance Condition
- Before Shipping to Service Center
- Data Backup and Recovery
- B1500A System Recovery
- Updating EasyEXPERT
- Error Codes
- 10 Application Library and Utilities

9-58 Agilent B1500 User’s Guide, Edition 7
If You Have a Problem
Error Codes
The time stamp function is not available for the binary data output
format. To use the time stamp function, set the data output format to
ASCII.
655 Cannot connect/disconnect series resistor.
The series resistor status cannot be changed in the high voltage state.
Set the output
voltage or the voltage compliance to ±42 V or less to
connect or disconnect the series resistor.
656 Series resistor must be OFF for 1 A range.
The series resistor cannot be set to ON for the measurement channels o
r
the output channels that use 1 A range.
657 Series resistor cannot be used with ASU.
The series resistor is not available for the channel connected
to the Atto
Sense and Switch Unit (ASU).
660 Sampling measurement was aborted.
Sampling measurement was aborted by the automatic abort function.
661 Negative hold time is only valid for I/V-t linear sampling with interval
< 2
ms.
662 Samplin
g interval for I/V-t log sampling must be 2 ms or longer.
663 Number of samples does not have to exceed 100001.
664 Base and bias must be same polarity for I mode.
The ba
se and bias values of the MI
command must be the same
polarity.
670 Specified channel does not have ASU.
Specify the module that can be used with the ASU.
671 SSP is not available for this channel.
SSP command is available only for the CMU. Specify the slot number
th
at the CMU ha
s been installed.
680 CMU correction mode must be manual.
To perform the CMU correction by using the ADJ? command, set the
CMU correct
ion mode to manual by using the ADJ command.
681 CMU correction mode must be off.
682 Invalid standard is specified as CMU correction.