User`s guide
Table Of Contents
- User’s Guide
- 1 Getting Started
- 2 Introduction
- 3 Installation
- 4 Using EasyEXPERT
- 5 Classic Test Definition
- I/V Sweep
- Multi Channel I/V Sweep
- I/V List Sweep
- I/V-t Sampling
- C-V Sweep
- Direct Control
- Function Setup
- Auto Analysis Setup
- Display Setup
- SMU Range Setup Window
- ADC and Integration Time Setup Window
- Advanced Setup Window
- CMU Range Setup Window
- Advanced Setup Window for C-V Sweep
- Switching Matrix Control
- SPGU Control
- SPGU Pulse Setup Window
- Load Z Setup Window
- Pulse Switch Setup Window
- SPGU ALWG Setup Window
- Define ALWG Waveform Window
- 6 Application Test Definition
- 7 Function Details
- I/V Sweep Measurement
- Multi Channel I/V Sweep Measurement
- I/V-t Sampling Measurement
- C-V Sweep Measurement
- SPGU Module
- Sweep Abort Function
- Standby Function
- Bias Hold Function
- Current Offset Cancel
- SMU CMU Unify Unit
- Atto Sense and Switch Unit
- SMU/PG Selector
- SMU Ranging Mode
- SMU Compliance
- SMU Pulse
- SMU Measurement Time
- SMU Filter
- SMU Series Resistor
- Interlock Function
- Auto Power Off Function
- Initial Settings
- 8 Built-in Programming Tool
- 9 If You Have a Problem
- When You Operate B1500A
- When You Perform Measurement
- Measurement Takes More Time than Specified
- Noise Affects the Measured Values
- Voltage Measurement Error is Large
- SMU Oscillates for High-Frequency Device Measurements
- SMU Oscillates for Negative Resistance Measurements
- Large Current Causes High Temperature (Thermal Drift)
- Measurement Damages the Device under Test
- Leaving Connections Damages Devices after Measurement
- Unexpected Sampling Measurement Data is Returned
- MFCMU Causes Unbalance Condition
- Before Shipping to Service Center
- Data Backup and Recovery
- B1500A System Recovery
- Updating EasyEXPERT
- Error Codes
- 10 Application Library and Utilities

Agilent B1500 User’s Guide, Edition 7 9-55
If You Have a Problem
Error Codes
Before triggering the pulsed spot C measurement, send the PDCV
command to set the pulsed voltage source.
282 Send PWDCV to set CV pulse sweep source.
Before triggering the pulsed CV measurement, send the PWDCV
command to set the pu
lsed voltage sweep source.
283 Set linear sweep for MM20.
Only the linear sweep is available for the PWDCV comma
nd for the
pulsed CV measurement (MM20).
284 Improper setting of CMU frequency and pulse width.
Pulse width value is out of the range for the CMU output frequency. Set
both frequ
ency value and pulse width value properly.
290 Send WFC to set Cf sweep source.
Before triggering the Cf sweep measurement, send the WFC command
to set the frequency
sweep source (oscillator).
301 Line power failure.
303 Excess voltage in MPSMU.
Voltage that exceeds maximum voltage a
t the present current r
ange was
detected by a MPSMU. All output switches were set to OFF.
304 Ground unit abuse is detected.
305 Excess current in HPSMU.
Current that exceeds maximum current at the present voltage range was
d
etected by
a HPSMU. All output switches were set to OFF.
307 Unsupported module.
This module is not supported by this firmware revision. Until you
update the fi
rmware, use Agilent B1500 with this module removed.
310 Interlock open operation error. Initialized.
Initialization was automatically performed because the B1500 failed to
set its output
to 0 V when the interlock circuit was opened in the high
voltage condition. Any module may be defective. Perform self-test.
311 ASU control cable was connected/discon
nected.
The B1500 must be turned off when the Atto Sense and Switch Unit
(ASU) is connected/disconn
ected.