User`s guide
Table Of Contents
- User’s Guide
- 1 Getting Started
- 2 Introduction
- 3 Installation
- 4 Using EasyEXPERT
- 5 Classic Test Definition
- I/V Sweep
- Multi Channel I/V Sweep
- I/V List Sweep
- I/V-t Sampling
- C-V Sweep
- Direct Control
- Function Setup
- Auto Analysis Setup
- Display Setup
- SMU Range Setup Window
- ADC and Integration Time Setup Window
- Advanced Setup Window
- CMU Range Setup Window
- Advanced Setup Window for C-V Sweep
- Switching Matrix Control
- SPGU Control
- SPGU Pulse Setup Window
- Load Z Setup Window
- Pulse Switch Setup Window
- SPGU ALWG Setup Window
- Define ALWG Waveform Window
- 6 Application Test Definition
- 7 Function Details
- I/V Sweep Measurement
- Multi Channel I/V Sweep Measurement
- I/V-t Sampling Measurement
- C-V Sweep Measurement
- SPGU Module
- Sweep Abort Function
- Standby Function
- Bias Hold Function
- Current Offset Cancel
- SMU CMU Unify Unit
- Atto Sense and Switch Unit
- SMU/PG Selector
- SMU Ranging Mode
- SMU Compliance
- SMU Pulse
- SMU Measurement Time
- SMU Filter
- SMU Series Resistor
- Interlock Function
- Auto Power Off Function
- Initial Settings
- 8 Built-in Programming Tool
- 9 If You Have a Problem
- When You Operate B1500A
- When You Perform Measurement
- Measurement Takes More Time than Specified
- Noise Affects the Measured Values
- Voltage Measurement Error is Large
- SMU Oscillates for High-Frequency Device Measurements
- SMU Oscillates for Negative Resistance Measurements
- Large Current Causes High Temperature (Thermal Drift)
- Measurement Damages the Device under Test
- Leaving Connections Damages Devices after Measurement
- Unexpected Sampling Measurement Data is Returned
- MFCMU Causes Unbalance Condition
- Before Shipping to Service Center
- Data Backup and Recovery
- B1500A System Recovery
- Updating EasyEXPERT
- Error Codes
- 10 Application Library and Utilities

Agilent B1500 User’s Guide, Edition 7 9-25
If You Have a Problem
Error Codes
105031 The product of voltage source value and current compliance value of A
exceeds its output power range. Please set the current compliance value
between B A and C A.
105032 The product of current source value and voltage compliance value of A
exceeds its output po
wer range. Please set the voltage compliance value
between B V and C V.
105033 The polarity of peak value(s) and the base value of SMU current pulse
must be th
e same.
105034 The start value and stop value of SMU log sweep must be non-zero and
have same pol
arity for primary sweep source (and synchronous sweep
source).
105035 Invalid value (A) was specified fo
r SMU I/V Sweep stop condition.
Valid option(s): B
A is not available for the setting of the sweep abort condition. Select
one from the available ch
oices B. If you use a formula, the calculation
result must show the available value.
105036 The hold time for SMU I/V Sweep must be between A s and B s.
105037 The delay
time for SMU I/V Sweep must be between A s and B s.
1050
41 Invalid value (B) was specified for current
monitor ranging mode of A.
Valid option(s): C
105042 Invalid value (B) was specified fo
r current monitor range of A. Valid
option(s): C
105043 Invalid value (B) was specified fo
r current monitor ranging rule of A.
Valid option(s): C
105044 Invalid value (B) was specified for volt
age monitor ranging mode of A.
Valid option(s): C
105045 Invalid value (B) was specified fo
r voltage monitor range of A. Valid
option(s): C
105046 Invalid value (A) was specif
ied for the integral mode of
High-Resolution ADC. Valid option(s): B
105047 The coefficient for High Resolution ADC A integr
al mode must be
between B and C.
105048 Invalid value (A) was specif
ied for the averaging mode of High-Speed
ADC. Valid option(s): B