User`s guide
Table Of Contents
- User’s Guide
- 1 Getting Started
- 2 Introduction
- 3 Installation
- 4 Using EasyEXPERT
- 5 Classic Test Definition
- I/V Sweep
- Multi Channel I/V Sweep
- I/V List Sweep
- I/V-t Sampling
- C-V Sweep
- Direct Control
- Function Setup
- Auto Analysis Setup
- Display Setup
- SMU Range Setup Window
- ADC and Integration Time Setup Window
- Advanced Setup Window
- CMU Range Setup Window
- Advanced Setup Window for C-V Sweep
- Switching Matrix Control
- SPGU Control
- SPGU Pulse Setup Window
- Load Z Setup Window
- Pulse Switch Setup Window
- SPGU ALWG Setup Window
- Define ALWG Waveform Window
- 6 Application Test Definition
- 7 Function Details
- I/V Sweep Measurement
- Multi Channel I/V Sweep Measurement
- I/V-t Sampling Measurement
- C-V Sweep Measurement
- SPGU Module
- Sweep Abort Function
- Standby Function
- Bias Hold Function
- Current Offset Cancel
- SMU CMU Unify Unit
- Atto Sense and Switch Unit
- SMU/PG Selector
- SMU Ranging Mode
- SMU Compliance
- SMU Pulse
- SMU Measurement Time
- SMU Filter
- SMU Series Resistor
- Interlock Function
- Auto Power Off Function
- Initial Settings
- 8 Built-in Programming Tool
- 9 If You Have a Problem
- When You Operate B1500A
- When You Perform Measurement
- Measurement Takes More Time than Specified
- Noise Affects the Measured Values
- Voltage Measurement Error is Large
- SMU Oscillates for High-Frequency Device Measurements
- SMU Oscillates for Negative Resistance Measurements
- Large Current Causes High Temperature (Thermal Drift)
- Measurement Damages the Device under Test
- Leaving Connections Damages Devices after Measurement
- Unexpected Sampling Measurement Data is Returned
- MFCMU Causes Unbalance Condition
- Before Shipping to Service Center
- Data Backup and Recovery
- B1500A System Recovery
- Updating EasyEXPERT
- Error Codes
- 10 Application Library and Utilities

Agilent B1500 User’s Guide, Edition 7 7- 61
Function Details
SMU Measurement Time
Multiple Measurement Channels
To use multiple measurement channels, define multiple measurement parameters in
the Display Setup screen or the Function Setup screen. Then, measurement
sequence, time data, and measurement time for each measurement point are as
follows.
• Sequence of measurement execution
Measurement channels
perform measurement in t
he order shown below.
1. Measurement channels which use the high speed AD converter and the
FIXED measurement range start measurement simultaneously (parallel
measurement).
2. After the parallel measurement, the other measurement channels perform
measurement s
equentially. The order is top to bottom of the channels defined
in the SMU Range Setup screen.
As a special case, for the linear sampling measurement with Interval < 2 ms
, all
channels start measurement simultaneously.
• Time data
Only the measurement start time of the channel which s
tarts measurement at
first is recorded.
• Measurement time
Measurement time is the sum of the measurement tim
e by all channels.