User`s guide
Table Of Contents
- User’s Guide
- 1 Getting Started
- 2 Introduction
- 3 Installation
- 4 Using EasyEXPERT
- 5 Classic Test Definition
- I/V Sweep
- Multi Channel I/V Sweep
- I/V List Sweep
- I/V-t Sampling
- C-V Sweep
- Direct Control
- Function Setup
- Auto Analysis Setup
- Display Setup
- SMU Range Setup Window
- ADC and Integration Time Setup Window
- Advanced Setup Window
- CMU Range Setup Window
- Advanced Setup Window for C-V Sweep
- Switching Matrix Control
- SPGU Control
- SPGU Pulse Setup Window
- Load Z Setup Window
- Pulse Switch Setup Window
- SPGU ALWG Setup Window
- Define ALWG Waveform Window
- 6 Application Test Definition
- 7 Function Details
- I/V Sweep Measurement
- Multi Channel I/V Sweep Measurement
- I/V-t Sampling Measurement
- C-V Sweep Measurement
- SPGU Module
- Sweep Abort Function
- Standby Function
- Bias Hold Function
- Current Offset Cancel
- SMU CMU Unify Unit
- Atto Sense and Switch Unit
- SMU/PG Selector
- SMU Ranging Mode
- SMU Compliance
- SMU Pulse
- SMU Measurement Time
- SMU Filter
- SMU Series Resistor
- Interlock Function
- Auto Power Off Function
- Initial Settings
- 8 Built-in Programming Tool
- 9 If You Have a Problem
- When You Operate B1500A
- When You Perform Measurement
- Measurement Takes More Time than Specified
- Noise Affects the Measured Values
- Voltage Measurement Error is Large
- SMU Oscillates for High-Frequency Device Measurements
- SMU Oscillates for Negative Resistance Measurements
- Large Current Causes High Temperature (Thermal Drift)
- Measurement Damages the Device under Test
- Leaving Connections Damages Devices after Measurement
- Unexpected Sampling Measurement Data is Returned
- MFCMU Causes Unbalance Condition
- Before Shipping to Service Center
- Data Backup and Recovery
- B1500A System Recovery
- Updating EasyEXPERT
- Error Codes
- 10 Application Library and Utilities

7- 36 Agilent B1500 User’s Guide, Edition 7
Function Details
Sweep Abort Function
Sweep Abort Function
The automatic sweep abort function is used to stop sweep measurement (increasing
or decreasing source output value) when one of the following conditions occurs.
This function is useful to reduce sweep time and to prevent damage to the device
during measurement.
• The output reaches voltage compliance or current compliance
• A measurement value exceeds the specified measur
ement range
• An SMU oscillates
• An error is caused in the MFCMU
The automatic abort function is enabled by using the Sweep [xxxx] status field on
th
e Measurem
ent Setup screen of the Classic Test definition.
• Sweep [CONTINUE AT ANY] status: Sets the abort function OFF.
• Sweep [STOP AT ANY ABNORMAL] status: Sets the abort function ON.
This function is available for the following measurement modes:
• “I/V Sweep Measurement”
• “C-V Sweep Measurement”
When abort occurs
After measurement is aborted, the source forces the following value.
• Sweep source: Start value
• Pulsed source or Pulsed sweep source: Pulse base value