User`s guide
Table Of Contents
- User’s Guide
- 1 Getting Started
- 2 Introduction
- 3 Installation
- 4 Using EasyEXPERT
- 5 Classic Test Definition
- I/V Sweep
- Multi Channel I/V Sweep
- I/V List Sweep
- I/V-t Sampling
- C-V Sweep
- Direct Control
- Function Setup
- Auto Analysis Setup
- Display Setup
- SMU Range Setup Window
- ADC and Integration Time Setup Window
- Advanced Setup Window
- CMU Range Setup Window
- Advanced Setup Window for C-V Sweep
- Switching Matrix Control
- SPGU Control
- SPGU Pulse Setup Window
- Load Z Setup Window
- Pulse Switch Setup Window
- SPGU ALWG Setup Window
- Define ALWG Waveform Window
- 6 Application Test Definition
- 7 Function Details
- I/V Sweep Measurement
- Multi Channel I/V Sweep Measurement
- I/V-t Sampling Measurement
- C-V Sweep Measurement
- SPGU Module
- Sweep Abort Function
- Standby Function
- Bias Hold Function
- Current Offset Cancel
- SMU CMU Unify Unit
- Atto Sense and Switch Unit
- SMU/PG Selector
- SMU Ranging Mode
- SMU Compliance
- SMU Pulse
- SMU Measurement Time
- SMU Filter
- SMU Series Resistor
- Interlock Function
- Auto Power Off Function
- Initial Settings
- 8 Built-in Programming Tool
- 9 If You Have a Problem
- When You Operate B1500A
- When You Perform Measurement
- Measurement Takes More Time than Specified
- Noise Affects the Measured Values
- Voltage Measurement Error is Large
- SMU Oscillates for High-Frequency Device Measurements
- SMU Oscillates for Negative Resistance Measurements
- Large Current Causes High Temperature (Thermal Drift)
- Measurement Damages the Device under Test
- Leaving Connections Damages Devices after Measurement
- Unexpected Sampling Measurement Data is Returned
- MFCMU Causes Unbalance Condition
- Before Shipping to Service Center
- Data Backup and Recovery
- B1500A System Recovery
- Updating EasyEXPERT
- Error Codes
- 10 Application Library and Utilities

4- 48 Agilent B1500 User’s Guide, Edition 7
Using EasyEXPERT
Switching Matrix Operation Panel
Switching Matrix Operation Panel
The Switching Matrix Operation Panel is opened by clicking the Switching Matrix
button, and is used to create the switch setup for Agilent B2200A/B2201A/E5250A
switching matrix, send the setup to the switching matrix, and so on. The switch
setup can be saved to the switch setup preset group. This group is different from the
preset group for test setups.
Switch Setup This area is used to create the switching matrix switch setup. You can create the
setu
p by click
ing the Input-Output cross point. The switch setup can be sent to the
switching matrix by clicking the Apply Switch Setup button.
Setup Name Name of switch setup
Clear All Sets all cross point open.
Input Shows the switching matrix input port and the measurement
resource conn
ected to it. This information is set in the
“Configuration”.
Label Shows the label set to the measurement resource.
The label is
set in the application test definition or the classic test setup.
ALL Checking this box sets all of this line to close.
Unchecking this box sets all of this line to open.
This box is gray and have a check when a cross point in this line
is s
e
t to
clo
s
e.
O
utput Output channels 1 to 12, 24, 36, or 48. Available numbers
depend on th
e number of the matrix cards installed in the
switching matrix. The field below the channel number is used to
enter the label for the output channel.
Export to Current
M
y Favori
te Group
This button converts the switch setup to the switching matrix setup data format of
the classic test, and saves it to the present My Favorite Setup group.
Color of cross point Switch setup
green close
gray open